Specifying Optical Filters

As the demand for high-precision optical filters continues to rise, so does the need to drive down the cost of end products. While such filters have especially rigorous coating requirements, they are often over-specified for the intended application, adding unnecessary complexity to the optical design and manufacturing process and increasing the final product cost. By determining the application’s actual needs and exploring alternatives, instrument or optical system developers can meet high-performance demands at a competitive price.

The market for high-performance diagnostic and biomedical instruments is on the rise, with attendant increases in cutting-edge research techniques and the demand for instruments that incorporate internal optical systems and/or components. This growth in demand, in turn, propels competition among optical manufacturers and sets high customer expectations in terms of instrument cost and performance.

Among the most sought-after optical elements fuelling this competition are biomedical filters. The industry uses many filter types in biomedical applications such as fluorescence microscopy, flow cytometry and Raman spectroscopy. Fluorescence microscopy, for instance, uses high-precision Bandpass Filters in conjunction with a 45° dichroic Beamsplitter to provide discrete excitation/emission wavelength separation while maximising the signal-to-noise ratio in the fluorescence image.

A Raman spectroscopy system, on the other hand, may employ a laser-line notch filter to remove spurious background from the probe laser, with a Long Pass Ultra-sharp Edge Filter to deeply attenuate Rayleigh scattering.

Given the accuracy required in such biotech applications, the filters used must perform to very demanding specifications. It’s not unusual for a fluorescence filter to require blocking better than OD6, transmission levels in the passband higher than 90% and edge steepness (defined here as wavelength separation between 50% transmission and OD >4) <0.5%.

Manufacturing a filter to such performance specifications requires a viable design and a high-quality coating process. A complex coating will expose even slight manufacturing flaws, so developers must consider several key design aspects prior to filter fabrication.

Biomedical filter characteristics

One key aspect of biomedical filters is their inherently thick coatings, resulting directly from the tight cut-on/cut-off tolerances and broadband coverage they are required to achieve. For instance, the coating thickness on a bandpass filter in the visible could range from 8 μm to more than 30 μm per face, depending on the blocking level and extent.

Another key aspect is coating density. A porous film will exhibit spectral shift when exposed to varying thermal conditions or changes in humidity (i.e., when exiting a vacuum pressure chamber and interacting with the atmosphere). Given that biomedical filters need center wavelength and full width half maximum tolerances on the scale of ±2 nm, it does not take much spectral shifting to result in a nonconforming filter. The coating thus must be dense enough that it incurs minimal to zero spectral shift over time.

Not every deposition process is well suited to producing coatings that meet such demanding spectral requirements. The technology must have the ability to control thickness deposition accurately to produce a robust coating with a minimal shift. Few coating technologies offer the needed capabilities while maintaining a cost-effective end price. Delta Optical Thin Film has found that only Advanced Plasma Source (APS) technology and magnetron sputtering technology can efficiently produce these demanding coating designs.

Cost metrics

When designing and specifying components for an optical system, the cost is always a critical factor. Biomedical filters cost more than their lower-precision equivalents because achieving high performance entails additional design and manufacturing processes.

Performance differences can even increment cost across a similar biomedical family, such as a Bandpass Filter with OD4 attenuation versus one with OD6. The only difference in the filters is that one is higher performing in terms of blocking depth and passband transmission and thus has a thicker coating, requiring more machine time to fabricate.

To save money, developers should perform a system analysis to determine the blocking they actually need. Of course, there are situations where the system absolutely requires extensive blocking to operate effectively, but too frequently developers over-specify attenuation because they don’t associate deeper/broader blocking with a substantial increase in cost.

Optical material specification

To save money, it is also useful to understand the cost drivers involved in filter specification. When requesting a custom filter quote, for example, the requirements you place on the optical substrate can greatly affect lead time and part cost. Here are some key considerations when specifying the material for a filter:

Substrate material

It is critical to use a substrate appropriate to the application wavelength. For high transmission in the ultraviolet, choose fused silica or an equivalent. Conversely, for transmission in the infrared, consider a material such as germanium or zinc sulfide.

It’s also important to know where the glass is transmitting and where extreme-absorption bands occur. The transmission range is highly dependent on the glass index because the index determines percentage loss due to Fresnel reflections. Most biomedical filters need a high and flat transmission for the intended application range.

To reduce cost and material lead time, though, avoid obscure and specialty glasses when a BK7 or Borofloat will do the trick. High-grade fused silica, for instance, costs on average 10 times more than a glass like Borofloat33. Similarly, coloured glasses, which can help provide deep and/or extended blocking, are often expensive and must be specialordered.

Another selection factor for filters used in fluorescence applications is the substrate’s autofluorescence. Substrate auto-fluorescence that is unaccounted for at the application wavelength could debilitate the system.

Workarounds exist, such as adding blocking after the glass to attenuate fluorescence, but the best approach is to minimise auto-fluorescence through careful substrate selection. Data shows that most fused silica glasses are good choices, with select BK7 materials close behind.

Surface and cosmetic quality

The thick coatings common to biomedical filters amplify any surface imperfections in the glass. Hence, demanding surface quality specifications are often difficult to meet and have a significant effect on post-coat yields. Meeting stringent quality requirements often makes it necessary to use higher-grade substrates, which can substantially increase costs.

Customers are often put off by minor blemishes, but it is important to realize that cosmetic imperfections do not necessarily hinder optical performance. A more expensive substrate might produce a more visually appealing part, but the same coating on a substrate with minor cosmetic flaws may offer the same optical performance for less. If cost reduction is a primary motivating factor, consider looser scratch-dig requirements.

Substrate thickness and shape

Unless the optical system in question has a unique configuration, use standard off-the-shelf component sizes whenever possible. Most optics vendors stock standard sizes, whereas fabricating a custom size entails a longer lead time and a higher rate per piece.

It’s advisable to research size availability early in the design phase to avoid having to order all custom optics to implement into your system. Economies of scale that exist from a manufacturing, sourcing, coating tooling and test setup perspective will most likely make a 25-mm standard size significantly more affordable than a 24.7-mm custom size.

Surface accuracy

Biomedical filters’ thicker, higher-stress coatings can compromise substrate surface flatness, so specify only the flatness really needed. Requiring quarter-wave accuracy can introduce additional manufacturing steps and might require the starting substrate to be eighth-wave or better to achieve quarter-wave after coating. A tight surface-accuracy specification also requires surface measurement after coating and often means adding SiO2 on the alternate face to compensate for warping.

Understanding the critical specifications in your design will help you immediately recognize and prepare for these cost drivers. In some cases, a tight specification may be imperative to the success of the project. In others, tight specifications serve only to deplete the budget unnecessarily because a looser specification could have been used to achieve the same results. Knowing which is which is essential to meeting both cost and performance goals.

One way to proceed is to work closely with a supplier starting early in the design phase. The vendor can help developers who are unfamiliar with the coating process or are unsure of how to specify a filter to meet system needs. Vendors can also help translate application needs into coating requirements and work with developers through several design iterations to optimize optical specifications for both cost and performance. Look for a vendor that can provide the following:

  • Wide variety in off-the-shelf items to facilitate rapid prototyping and design iterations
  • Fast turnaround and feedback on custom design efforts
  • Excellent technical support and information online, such as filter transmission and blocking curves
  • Ready OEM support to allow a fast transition from prototypes to cost-effective production volumes without requiring design changes

With such vendor support, developers are well-positioned to take maximum advantage of costreduction opportunities in biomedical instrument design. By avoiding over-specification, developers can contain costs without affecting system performance.